Tri Tech Medical Gas Manifold

  • Pic
  • Main Pic
  • Capture
The Tri-Tech Medical fully automatic manfiold offers a turn-key solution to add or replace a medical gas manifold system.

Features
  • Complete system cleaned for use with Oxygen
  • Adjustable single stage line regulator 0 - 125 and 0 – 200 psig for Nitrogen
  • Includes copper or stainless steel cylinder pigtails
  • Outlet connection – 1/4 Male NPT
  • Wall mounting bracket included
  • Maximum inlet pressure 3,000 psig
  • Optional remote audio visual alarm
  • Five year warranty
 *WHEN REQUESTING A QUOTE PLEASE LET US KNOW HOW MUCH SPACE YOU HAVE TO WORK WITH AND HOW MANY TANKS ARE REQUIRED ON EACH SIDE.  CUSTOM CONFIGURATIONS ARE AVAILABLE

 

Specification
The NFPA 99 compliant digital, fully automatic manifold shall be a Tri-Tech Medical Genesys CC series. No
manual resetting of valves or levers shall be required. The unit shall switch from “Bank in Use” to “Reserve” bank without fluctuation in line delivery pressure.
Simultaneously, the “Reserve in Use” alarm shall be triggered by the manifold microprocessor. The manifold
shall continue to provide gas, in the event of a power failure, until both banks are depleted. After the switchover, the
“Reserve” bank shall then become the “Bank in Use”. The manifold microprocessor shall also trigger the “High Line
Pressure” and “Low Line Pressure” alarms without the need for additional pressure switches or transducers. The manifold shall be capable of being upgraded after installation, to be used with low or medium pressure portable bulk vessels or for use at higher or lower delivery pressures.
The microprocessor based control panel shall incorporate LED’s and an illuminated text display and shall provide
electronic monitoring of circuits with up to 20 error, alarm or information messages displayed for ease of maintenance.
The illuminated text display shall be readable even in poor lighting conditions. Analog gauges are also provided so that
line and both bank pressures may be observed in the event of a power failure. The control panel shall also incorporate a
set of LED’s for each bank, green for “Bank in Use”, amber for “Ready” and red for “Empty”.
All manifold regulators, piping and control switching equipment shall be cleaned for use with oxygen service and
installed in a steel powder coated cabinet (weatherproof version available) to provide protection and minimize tampering.